Overview
-
Tritium Column is designed as an alternative or compliment to distillation for most routine tritium analyses of aqueous samples. The Tritium Column works by removing metal ions which may be potential interferences in the LSC spectrum of 3H. The Tritium Column is not intended to be an enrichment procedure. Therefore, the Tritium Column should only be used when the required detection limit can be achieved by the direct counting of a 5-10 mL aliquot of the sample processed through the tritium column.
The Diphonix® Resin removes cations in exchange for hydrogen ions, with a theoretical capacity is 0.8 meq per Tritium Column. The anion exchange resin, a chloride form analytical grade 1X8 resin, exchanges anions in the sample for chloride ions. The polymethacrylate resin removes organically bound tritium and carbon-14 and other organic material which may cause quenching in the LSC cocktail.Please contact us at for specific academic pricing.
-
- Properties
- Applications
-
Overview
Figure 1 shows the composition of the Tritium Column.
Table 1 explains the purpose and capacity of each component.
Table 2 shows the average spike recoveries for a variety of sample matrix types passed through the Tritium Column.
The column successfully removes potential radioactive interferences. In a study, a mixture of radioisotopes (60Co, 137Cs, 233U, 90Sr/90Y, 210Pb, 230Th, total activity 16.9 Bq) was spiked into eight tritium-spiked solutions (distilled and sea water samples). In every case, after passing the sample through the Tritium Column, the number of counts in the region above the tritium window was not different than the number of counts in the same region of an unspiked sample.
Tritium Column
Tritium Column
Catalog Number:
SPE1556047EIC
Mfr. No.:
H3-C50-A
Price:
$1,750
Size:
50 Columns (2ml, 75-150 μm)