Overview
-
The Griess assay is a simple and popular method for detecting NO concentration. 2,3-Diaminonaphthalene (DAN) is a highly sensitive alternative to the Griess assay. The DAN method is 50-100 times more sensitive than the Griess assay: While the detection limit of the Griess assay is 1 mM, the limit of the DAN method is 10-50 nM. DAN reacts with NO2- in acidic conditions to produce fluorescent naphthalenetriazole. The wavelength of the emission maximum of naphthalenetriazole is 410 nm. However, detection at 450 nm is recommended to avoid fluorescent blanks and increase sensitivity. The fluorescent background of DAN is low for maximum sensitivity. The optimal reaction conditions of DAN with NO2- have been determined. The reaction should proceed at pH 2 at room temperature for 5 minutes, and the resulting fluorescence of naphthalenetriazole should be determined at a pH of 10 or more. DAN is a photosensitive reagent and sometimes becomes dark brown colored crystals. Since this brown product cannot be utilized for the fluorescent detection, recrystallization is necessary.
Please contact us at for specific academic pricing.
-
- Properties
- Applications
- Reference
-
Overview