Introduction
-
BJP 10 benchtop devices allow concentration of clinical samples up to 10 ml in volume. They can operate without a centrifuge and are intended for in vitro diagnostic (IVD) applications. The dead stop design provides maximum concentrate recovery where samples can be withdrawn without the need for glass Pasteur pipettes. BJP 10 devices have been used extensively to concentrate urine and spinal fluid samples prior to gel and immunofixation electrophoresis for diagnosis of multiple myeloma, light chain amyloidosis and related disorders. Savings of 30% have been reported when switching to BJP 10 from similar ultrafiltration devices.
Up to 10 ml starting volume
Concentrate up to 200x
• Applications
Concentration of urine prior to electrophoresis for diagnosis of multiple myeloma and amyloidosis.
Concentrate spinal fluid prior to electrophoresis for diagnosis of meningitis and multiple sclerosis.
Increase titers of bacterial antibodies (H. influenzae, S. pneumoniae, Streptococcus) prior to latex agglutination.
Simple concentration of biological samples and dilute protein solutions.
• Performance
Time to concentrate[minutes]at 20°C and solute recovery% (Different conc.factors shown)
Start Volume 10 ml(10x conc.) 10 ml(50x conc.) Min. Rec. Min. Rec. BSA 1.0 mg/ml(66,000 MW) 7,500 MWCO PES 100 92% 210 90% IgG 0.25 mg/ml(160,000 MW) 7,500 MWCO PES 110 78% 130 74% Technical Specifications
BJP-10/40 BJP-10/40P
BJP-10 (others)
Concentrator Capacity Per Sample 10 ml 10 ml Dimensions Total Length 147 mm 38 mm Width 70 mm 20 mm Height 94 mm 100 mm Active Membrane Area 25 cm² 25 cm² Dead stop volume 50µl 50µl Materials of Construction Reservoir Material Styrene-acrylonitrile Styrene-acrylonitrile Membrane Polyethersulfone Polyethersulfone
-
- Certificate
-
Introduction
-
ManualSafety Data Sheets (SDS)