                                            Overview
                                            - 
                                                    • In gradient elution analysis, impurities from the mobile phase solvents (such as mobile phase containers, flow lines, pumps, gradient mixer, etc.) can lead to the appearance of ghost peaks in chromatograms. 
 • The Ghost Clear Column is designed to adsorb and remove traces of impurities from the mobile phase, thereby suppressing the appearance of ghost peaks.
 • The Ghost Clear Column should be installed between Mixer and Injector.
 • Installing the Ghost Clear Column after injector would lead to strong adsorption of samples and affect analysis.
 • Not all impurities can be adsorbed by the Ghost Clear Column.
 • For new columns, flush the Ghost Clear Column with 80% methanol solution at 1 mL/min for 30 min before new column switching to the system.
 • Ion-pair solvents in mobile phase would be retained by the Ghost Clear Column and affect retention and peak shape.
 • Column lifetime depends on analytical conditions, mobile phase, solvent purity, etc.
 • Routine change of the Ghost Clear Column is suggested to ensure performance.Please contact us at for specific academic pricing. 
 
- 
                                                    