VerSus Reactor for Animal Cell Culture
| Cat. No. | Specification | Unit | Price | Quantity |
|---|---|---|---|---|
| BAF1552100-01 | Culture operation volume: 0.3L | Each | Inquiry |
|
| BAF1552100-02 | Culture operation volume: 1.2L | Each | Inquiry |
|
| BAF1552100-03 | Culture operation volume: 2.4L | Each | Inquiry |
|
| BAF1552100-04 | Culture operation volume: 7L | Each | Inquiry |
|
Product Description
Product Description
The VerSus Reactor is a new and innovative BioReactor for animal cell culture in which the technology of “SPG membrane sparger*” by which micro-bubbles with extremely high homogeneity can be generated is fused with a VMF Reactor. Through an efficient oxygen supply by the SPG membrane sparger, oxygen can be supplied efficiently without putting stress on the animal cells in the culture tank. Also, due to a decline in the DO air flow rate, a foam layer can be prevented. With the help of the dedicated controller “S-BOX × 10α II”, changes in various parameters including PI control can be easily performed.
Specifications
| Item | Specifications | ||||
|---|---|---|---|---|---|
| Name | VerSus Reactor | ||||
| Model | VSR-05 | VSR-1 | VSR-3 | VSR-10 | |
| Temp. control device | Band heater (PID control) with over-temperature protection function (Max. 80 [°C]) | ||||
| Power transmission | Linear shaft drive non-sealing type | ||||
| Gas supply type | Shirasu porous glass (SPG) membrane type (additional option : sintered metal type) | ||||
| Performance*1 | Temp. control range | Room temperature + 5 - 20 [°C] (normally set to 37 [°C]) | |||
| Temp. accuracy | ±0.3 [°C] (37 [°C]) | ||||
| Max. shaft stroke | 40 [mm] | ||||
| Max. shaft speed | 300 [mm/s] | ||||
| Function | Temp. setting | Touch panel input, data output DC1 - 5 [V] | |||
| Vertical motion setting | Touch panel input, data output DC0 - 5 [V] | ||||
| Configuration | Power of band heater | 60 [W] | 105 [W] | 160 [W] | 480 [W] |
| Power of motor | Max. power 800 [W] | ||||
| Mixing impellers | VM200 | VM100+VM200*2 | |||
| Culture tank | Dimensions | I.D. 90 x Depth 200 [mm] | I.D. 110 x Depth 169 [mm] | I.D. 140 x Depth 203 [mm] | I.D. 200 x Depth 360 [mm] |
| Culture operation volume | 0.3 [L] | 1.2 [L] | 2.4 [L] | 7 [L] | |
| Usage condition | 10 - 35 [°C] | ||||
| Outer dimensions | W360×D485×H905 [mm] | ||||
| Weight | Approx. 28 [kg] | Approx. 28 [kg] | Approx. 30 [kg] | Approx. 34 [kg] | |
| Power supply | AC100 [V], 50/60 [Hz] | ||||
| *1 The performance is the value under room temperature 20 [°C], power supply AC100 [V], 50 [Hz] and no load condition. | |||||
| *2 Depending on the Culture operation volume, the only mixing impeller may be only the VM200. | |||||
| Item | Specifications | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Name | Culture controller | |||||||||||
| Model | S-BOX×10αII | |||||||||||
| Control | pH, DO (Dissolved oxygen), FL (O₂ flow rate), Pump | |||||||||||
| Display range/accuracy | pH (Hydrogen ion concentration) | 0.00 - 14.00 [-] | Display accuracy: ±0.5 [%] F.S. | |||||||||
| DO (Dissolved oxygen) | 0.00 - 20.00 [mg/L] | |||||||||||
| FL (O2flow rate) | 0.00 - 20.00 [mL/min] | |||||||||||
| Setting range | pH (Hydrogen ion concentration) | 0.00 - 14.00 [-] | ||||||||||
| DO (Dissolved oxygen) | 0.00 - 10.00 [mg/L] | |||||||||||
| FL (O2flow rate) | 0.00 - 20.00 [mL/min] | |||||||||||
| Control type | pH (Hydrogen ion concentration) | ON/OFF control | ||||||||||
| DO (Dissolved oxygen) | O2addition/subtraction step control | |||||||||||
| O2and N2ON/OFF control | ||||||||||||
| O2PI control (slope set method: TIME, %) | ||||||||||||
| FL (O2flow rate) | PI control (slope set method: TIME, %) | |||||||||||
| Data output | pH (Hydrogen ion concentration) | With data logger DC0 - 5 [V] (option : Load factor can be outputted.) |
Accuracy: ±0.5 [%] F.S. | |||||||||
| DO (Dissolved oxygen) | ||||||||||||
| FL (O2flow rate) | ||||||||||||
| Mixing frequency | ||||||||||||
| Temperature sensor | DC1 - 5 [V] | |||||||||||
| MTA of outer surface | SUS304 (no coating), indoor type, non-waterproof, non-explosion proof specification | |||||||||||
| Installation | Indoor tabletop type | |||||||||||
| Outer dimensions/weight | W350×D400×H530 [mm] ・ Approx. 15 [kg] | |||||||||||
| Usage conditions | Temperature | 5 - 45 [°C] | Humidity | 20 - 85 [%] RH (No condensation) | ||||||||
| Sensors | Polarographic DO sensor/pH sensor manufactured by Mettler Toledo company | |||||||||||
| (Option: Optical DO sensor manufactured by Automatic System Research Co., Ltd. ) | ||||||||||||
| Utilities | Power supply | AC100 [V], 50/60 [Hz], electrical outlet 2 gang (for main control unit and recorder) | ||||||||||
| N2 | Flow rate 50 [mL/min] or lower, supply pressure 0.2 [MPa], connection port φ6 one touch tube fitting | |||||||||||
| O2 | Flow rate 20 [mL/min] or lower, supply pressure 0.2 [Mpa] connection port φ6 one touch tube fitting | |||||||||||
| CO2 | Flow rate 50 [mL/min] or lower, supply pressure 0.2 [MPa], connection port φ6 one touch tube fitting | |||||||||||
| AIR | Flow rate 150 [mL/min] or lower, supply pressure 0.2 [MPa], connection port φ6 one touch tube fitting | |||||||||||
| *All of N2, O₂, CO₂, and AIR must be dry and clean gases that do not contain corrosive components, dust, and oil mist. | ||||||||||||