VerSus Reactor for Animal Cell Culture

Cat. No. Specification Unit Price Quantity
BAF1552100-01 Culture operation volume: 0.3L Each Inquiry
BAF1552100-02 Culture operation volume: 1.2L Each Inquiry
BAF1552100-03 Culture operation volume: 2.4L Each Inquiry
BAF1552100-04 Culture operation volume: 7L Each Inquiry

Add to Cart

Product Description
Product Description
The VerSus Reactor is a new and innovative BioReactor for animal cell culture in which the technology of “SPG membrane sparger*” by which micro-bubbles with extremely high homogeneity can be generated is fused with a VMF Reactor. Through an efficient oxygen supply by the SPG membrane sparger, oxygen can be supplied efficiently without putting stress on the animal cells in the culture tank. Also, due to a decline in the DO air flow rate, a foam layer can be prevented. With the help of the dedicated controller “S-BOX × 10α II”, changes in various parameters including PI control can be easily performed.
Specifications
Item Specifications
Name VerSus Reactor
Model VSR-05 VSR-1 VSR-3 VSR-10
Temp. control device Band heater (PID control) with over-temperature protection function (Max. 80 [°C])
Power transmission Linear shaft drive non-sealing type
Gas supply type Shirasu porous glass (SPG) membrane type (additional option : sintered metal type)
Performance*1 Temp. control range Room temperature + 5 - 20 [°C] (normally set to 37 [°C])
Temp. accuracy ±0.3 [°C] (37 [°C])
Max. shaft stroke 40 [mm]
Max. shaft speed 300 [mm/s]
Function Temp. setting Touch panel input, data output DC1 - 5 [V]
Vertical motion setting Touch panel input, data output DC0 - 5 [V]
Configuration Power of band heater 60 [W] 105 [W] 160 [W] 480 [W]
Power of motor Max. power 800 [W]
Mixing impellers VM200 VM100+VM200*2
Culture tank Dimensions I.D. 90 x Depth 200 [mm] I.D. 110 x Depth 169 [mm] I.D. 140 x Depth 203 [mm] I.D. 200 x Depth 360 [mm]
Culture operation volume 0.3 [L] 1.2 [L] 2.4 [L] 7 [L]
Usage condition 10 - 35 [°C]
Outer dimensions W360×D485×H905 [mm]
Weight Approx. 28 [kg] Approx. 28 [kg] Approx. 30 [kg] Approx. 34 [kg]
Power supply AC100 [V], 50/60 [Hz]
*1 The performance is the value under room temperature 20 [°C], power supply AC100 [V], 50 [Hz] and no load condition.
*2 Depending on the Culture operation volume, the only mixing impeller may be only the VM200.
Item Specifications
Name Culture controller
Model S-BOX×10αII
Control pH, DO (Dissolved oxygen), FL (O₂ flow rate), Pump
Display range/accuracy pH (Hydrogen ion concentration) 0.00 - 14.00 [-] Display accuracy: ±0.5 [%] F.S.
DO (Dissolved oxygen) 0.00 - 20.00 [mg/L]
FL (O2flow rate) 0.00 - 20.00 [mL/min]
Setting range pH (Hydrogen ion concentration) 0.00 - 14.00 [-]
DO (Dissolved oxygen) 0.00 - 10.00 [mg/L]
FL (O2flow rate) 0.00 - 20.00 [mL/min]
Control type pH (Hydrogen ion concentration) ON/OFF control
DO (Dissolved oxygen) O2addition/subtraction step control
O2and N2ON/OFF control
O2PI control (slope set method: TIME, %)
FL (O2flow rate) PI control (slope set method: TIME, %)
Data output pH (Hydrogen ion concentration) With data logger DC0 - 5 [V]
(option : Load factor can be outputted.)
Accuracy: ±0.5 [%] F.S.
DO (Dissolved oxygen)
FL (O2flow rate)
Mixing frequency
Temperature sensor DC1 - 5 [V]
MTA of outer surface SUS304 (no coating), indoor type, non-waterproof, non-explosion proof specification
Installation Indoor tabletop type
Outer dimensions/weight W350×D400×H530 [mm] ・ Approx. 15 [kg]
Usage conditions Temperature 5 - 45 [°C] Humidity 20 - 85 [%] RH (No condensation)
Sensors Polarographic DO sensor/pH sensor manufactured by Mettler Toledo company
(Option: Optical DO sensor manufactured by Automatic System Research Co., Ltd. )
Utilities Power supply AC100 [V], 50/60 [Hz], electrical outlet 2 gang (for main control unit and recorder)
N2 Flow rate 50 [mL/min] or lower, supply pressure 0.2 [MPa], connection port φ6 one touch tube fitting
O2 Flow rate 20 [mL/min] or lower, supply pressure 0.2 [Mpa] connection port φ6 one touch tube fitting
CO2 Flow rate 50 [mL/min] or lower, supply pressure 0.2 [MPa], connection port φ6 one touch tube fitting
AIR Flow rate 150 [mL/min] or lower, supply pressure 0.2 [MPa], connection port φ6 one touch tube fitting
*All of N2, O₂, CO₂, and AIR must be dry and clean gases that do not contain corrosive components, dust, and oil mist.