6.2ft. Width 20" Opening Class II B2 Biological Safety Cabinet

Cat. No. Dimensions (WxDxH) Specification Package Unit Price Quantity
BC01503772-01 1700×600×660 mm ULPA Filter: Two, 99.9995% efficiency at 0.12 μm, filter life indicator. Main Body: 2000×950×1820 mm EA $4,440

Add to Cart

Product Description
Product Description
This products belong to Class II B2 biological safety cabinet which fully meet the requirement of US standard ANSI/NSF49:2002, European standard EN12469:2000, biological safety cabinet is a kind of negative pressure filtration system for protecting operator, the laboratory environment and work materials, the front opening which air flow inward have protection function for operator, the filtered laminar flow generated by vertical HEPA can protect work materials, what's more, the polluted air flow become pure after processed by HEPA(ULPA) filter. When it's used in microbiology experiment environment filled with volatile or toxic chemical and radionuclide, suitable exhaust hood in function have to be linked.
Three protection:operator, sample and environment.
Airflow system: 0% air recirculation, 100% air exhaust.
A Class Il B2 BSC, also called a total exhaust cabinet, is necessary when significant amounts of radionuclides and volatile chemicals are expected to be used.
Features
· Time reserve function.
· Motorized front window.
· HEPA filter life and UV life indicator.
· Automatic air speed adjustable with filter block.
· With memory function in case of power-failure.
· Work area surrounded by negative pressure, it can ensure maximum safety in work area.
· Most accessories are standard, which saves your money. There is no need to pay more.
· Audio and visual alarm (filter replacement, window over height, abnormal air flow velocity, etc).
· Remote control. All functions can be realized with it, making the operation much easier and more convenient.
· Interlock function: UV lamp and front window; UV lamp and blower, fluorescent lamp; blower and front window.
· Foot switch. Adjust front window height by foot during experiment, to avoid airflow turbulence caused by arm movement.
Product Details
Accessory

Items

Quantity

Main Body

1unit

Base Stand

1 set

External exhaust blower

1set

External exhaust blower holder

1set

Exhaust duct

1pc

Exhaust duct clip

2pc

Power cord 

1pc

Fuse(15A)

2pcs

Fuse(5A)

1pc

UV Lamp  (T6 40W)

1pc

Remote control (including battery)

1pc

Keys

2pc

Foot switch

1pc

User manual

1pc

Test report

1pc

Quality certification card

1pc

Drain Valve

1set

Inner hex cylinder head bolt M10×55

1pc

Inner hex cylinder head bolt M10×80+flat washer 10+spring washer10+M10 nut

4sets

Inner hex cylinder head bolt M10×25+flat washer 10+spring washer 10

4sets

Inner hex cylinder head bolt M8×25+flat washer 8+spring washer 8+ M8 nut

4sets

Inner hexagon wrench

1pc

Big rubber gasket

1pc

Small rubber gasket

1pc

Front window tubular motor control rod

1pc

M8*100 Expansion bolt

6pcs

Transparent glass glue (with rubber mouth)

1pc

Buckle stopper (white)

7pcs

Connecting fittings

4pcs

Specifications

Model

BSC-1800IIB2-X

Internal Size(W×D×H)

1700×600×660 mm

External Size(W×D×H)

1873×775×2270mm

Tested Opening

Safety Height 200 mm(8")

Max Opening

480mm(20")

Inflow Velocity

0.53±0.025 m/s

Down flow Velocity

0.33±0.025 m/s

Pre-Filter

Washable

ULPA Filter

Two, 99.9995% efficiency at 0.12 μm,filter life indicator.

Front Window

Motorized,two-layer laminated toughened glass ≥ 5mm,anti UV.

Noise

NSF49 ≤ 61 dB / EN1246949 ≤ 58 dB 

UV Lamp

40W×1

UV timer, UV life indicator, emission of 253.7 nanometers 

for most efficient decontamination.

Illuminating Lamp

LED Lamp

16W×2

Illumination

≥1000Lux

Consumption

1200W

Waterproof Sockets

Two, total load of two sockets: 500W

Display

LCD display: exhaust filter and downflow filter pressure, filter and UV lamp working time, inflow and downflow velocity, filter life, humidity and temperature, system working time etc.

Control System

Microprocessor

Airflow System

0% air recirculation, 100% air exhaust

Alarm

Abnormal airflow velocity; Filter replacement; Front window at unsafe height.

Exhaust Duct

4 meters PVC duct, Diameter:300mm

Material

Work Zone: 304 stainless steel

Main Body: Cold-rolled steel with anti-bacteria powder coating.

Work Surface Height

750mm

Caster

Foot master caster

Power Supply

AC 220V±10%, 50/60Hz; 110V±10%, 60Hz (110V/60Hz is not applicable to BSC-1800IIB2-X)

Standard Accessory

Illuminating lamp, UV lamp×2,base stand, remote control, foot switch, exhaust blower, 

exhaust duct, drain valve, waterproof sockets×2, pipe strap×2

Optional Accessory

Water and gas tap, Electric height adjustable base stand

Gross Weight

408kg

Package

Main Body

2000×950×1820 mm

Exhaust Blower(W×D×H)

970× 810× 680 mm