TYD02-01 Laboratory Syringe Pump, 0.184nL/min-173.718mL/min Flow, 1 Channel

Syringe size: 10uL-140mL
Work mode: infusion, withdraw, infusion/withdraw, withdraw/infusion, continuous mode
Channel number: 1
Stroke of pump: 140mm
Pusher advance per microstep: 0.156um/ustep
Linear speed: 1um/min-150mm/min
Linear resolution: 1um/min
Linear travel accuracy: error <± 0.35% (>30% of pump stroke);
Linear force: maximum rated: > 20KGF
Thrust regulation: 1-100% arbitrary adjustable
Syringe selection: built-in main manufacturers, the main model syringe to choose;Can use custom syringe, direct input syringe size & diameter
Flow calibration: through the calibration procedure to obtain more accurate fluid volume
Display: 4.3-inch color LCD, 480*272 resolution, color 65k; With screen display and transfusion volume, quantity of liquid, flow, operation direction, syringe specification, etc.; Animation display operation status
Operation mode: touch screen + button
Power-off memory: storing the running parameters automatically
Other features: pause and stop; The sound tip; The lock parameters; Fast forward and retreat quickly
State signal output: 1 road stop state, 1 state road direction
Control signal input: 1 way down along the priming, 1 way down along the trigger to stop
Communication interface: RS485, MODBUS protocol
Applicable power: AC100-240V/20W
working temperature: 5 to 40°C
Relative humidity: <80% RH
Dimension: 290x205x180 mm
Weight: 4.4kg
Cat. No. Dimensions (WxDxH) Specification Unit Price Quantity
LP01410133-01 290x205x180 mm Flow: 0.184nL/min-173.718mL/min; Max: 1 Channel EA Inquiry

Add to Cart

Product Description
Product Description
TYD02-01 laboratory syringe pump is a support for a variety of work mode of high precision syringe pump, high resolution color LCD, touch screen is convenient fast set parameters, multiple light is clear and confirmed its working status, and supports a variety of syringe to choose, to meet the requirements of different experimental flow and high precision control, and protection agencies and alarm mechanism. Support infusion/withdraw five kinds of working mode.Built-in RS-485 communication, support the MODBUS protocol. All-metal anti-static, anti-corrosion spraying process housing, wide range power input, adapt to various occasions. TYD02-01 suitable for syringe size 10uL-140mL; Linear force 1um/min-150mm/min.
Features
▪It can assemble one syringe.
▪A variety of working mode choice.
▪Color touch screen, LCD display, convenient and quick operation.
▪Support screen lock, key mute operation.
▪With lights, buttons, clearly working state.
▪Support a variety of standard syringe, customizable syringe.
▪High precision control.
▪Syringe protection and traffic jam alarm.
▪RS485 communication, support the Modbus protocol.
▪External signal control start-stop and direction.
▪A wide range of power input.
▪All-metal anti-static, anti-corrosion spraying process housing.
Specifications
Syringe size Syringe ID (mm) Min. flow rate (nl/min) Max .flow rate (ml/min)
10μl 0.485 0.184 0.027
25μl 0.729 0.417 0.0626
50μl 1.03 0.833 0.125
100μl 1.457 1.667 0.250
250μl 2.304 4.169 0.625
500μl 3.256 8.326 1.248
1ml 4.699 17.342 2.601
5ml 11.989 112.890 16.933
10ml 14.427 163.469 24.520
20ml 19.05 285.027 42.754
30ml 21.59 366.090 54.913
60ml 26.594 555.459 83.318
100mL 35.7 1000.982 150.147
140mL 38.4 1158.117 173.718