| Cat. No. | Dimensions (WxDxH) | Specification | Unit | Price | Quantity |
|---|---|---|---|---|---|
| OPI1578952-01 | 47” x 30.25” x 47” | Footprint: 12.5 sq ft; Bench Weigh: 560 lbs (max); Ship Weight: 650 Ibs (max); Noise Output: <70 dB | EA | Inquiry |
|
Product Description
Product Description
Our system is designed to meet high-volume testing demand. With continuous sample loading, the elimination of most manual liquid transfers, and reduced hands-on time by up to 60%, your lab can rely on this trusted workhorse.
Product Details
Software
Controlling Software: Agility® Software
Work Protocols (Assays): Unlimited
Data Processing: Quantitative and qualitative
Levey-Jennings: Yes
Westgard Rules: Yes
Process Reporting: Event log + error log
Automatic Error Recovery: Yes
Password Access Control: Yes
Work Protocols (Assays): Unlimited
Data Processing: Quantitative and qualitative
Levey-Jennings: Yes
Westgard Rules: Yes
Process Reporting: Event log + error log
Automatic Error Recovery: Yes
Password Access Control: Yes
Specifications
Supply Requirements
Voltage: 100 – 240 V automatic conversion
Frequency: 50/60 Hz
Power Consumption: < 400 VA (online UPS recommended)
System Specifications
Number of Plates: 12 (additional continuous load capability)
Sample Capacity: 200 (additional continuous load capability)
Continuous Sample Loading: Yes
Sample Tube Size: 10 – 17 mm diameter / 45 – 100 mm height
Number of Reagents: Up to 12 reagent packs simultaneously (additional packs can
be added using continuous load capability)
Number of Standard/Control Bottles: Included in each reagent pack
Assays per Plate: Up to 4
Dilution Wells Capacity: Up to 384 deep-well strips (48 x 8)
Sample Tip Capacity: Up to 2,240 tips (20 x 112 with additional continuous load capability)
Reagent Tip Capacity: Up to 98 tips (additional continuous load capability)
Reader Specifications
Photometric Range: 0.000 – 3.500 OD
Spectral Range: 405 – 690 nm
Filter Slots: 6 wavelength
Precision: ± 0.01 OD at 0.000 – 0.500 OD
< 1% CV (< 2.0 OD)
< 1.5% CV (2.0 – 2.5 OD)
Accuracy: ± 0.01 OD or 2.5%
(0.000 – 3.000 OD) whichever is greater
Read Time: < 30 sec (single wavelength)
< 50 sec (dual wavelength)
Washer Specifications
Manifold Configuration: 8-channel wash head
Programmable Volumes: 50 – 999 μL
Wash Buffer Capacity: 5 wash bottles at 2.0 L, with level-sensing
Residual Wash Volume: < 3 μL per well with dual-axis sweep in a flat-bottom plate
Dispense Precision: ≤ 5% CV (with 300 μL in a 96 well plate)
Waste Container Capacity: 10 L, with level-sensing
Incubator Specifications
Number of Incubators: 12 (6 ambient, 6 elevated)
Temperature Range: RT + 4° – 45°C (elevated incubator)
RT + 4°C (ambient incubator)
Temperature Accuracy: ± 1°C
Shaking: 14 Hz (periodic or continuous)
Temperature Monitoring: Yes
Reagent Pipetting Specifications
Reagent Tip Size: 1,200 μL
Reagent Pipetting Volume: 20 – 1,000 μL
Reagent Pipetting Precision: ≤ 3% CV at 10 shots at any volume in operating range
Reagent Pipetting Accuracy: ± 2% of target volume (single-shot mode)
Sample Pipetting Specifications
Sample Tip Size: 300 µL
Sample Pipetting Volume: 10 – 300 µL
Sample Pipetting Precision: ≤ 3% CV at any operating volume above 10 µL
Sample Pipetting Accuracy: ± 2% of target volume
Time to Dispense: 11 minutes (typical)
50 µL of 96 samples to plate from sample tubes or deep-well plates
Dilution Range: 1 part in 199 one-stage dilution
1 part in 39,601 two-stage dilution
Process Security
Liquid Level Sensing: Yes (reagents, controls and samples)
Level Sensor System Pressure differential
Clot Detection: Yes
Dispense Anomaly Detection: Yes
Tip Detection: Yes
Electronic Signature Pipetting: Yes
Alarms: Yes
Voltage: 100 – 240 V automatic conversion
Frequency: 50/60 Hz
Power Consumption: < 400 VA (online UPS recommended)
System Specifications
Number of Plates: 12 (additional continuous load capability)
Sample Capacity: 200 (additional continuous load capability)
Continuous Sample Loading: Yes
Sample Tube Size: 10 – 17 mm diameter / 45 – 100 mm height
Number of Reagents: Up to 12 reagent packs simultaneously (additional packs can
be added using continuous load capability)
Number of Standard/Control Bottles: Included in each reagent pack
Assays per Plate: Up to 4
Dilution Wells Capacity: Up to 384 deep-well strips (48 x 8)
Sample Tip Capacity: Up to 2,240 tips (20 x 112 with additional continuous load capability)
Reagent Tip Capacity: Up to 98 tips (additional continuous load capability)
Reader Specifications
Photometric Range: 0.000 – 3.500 OD
Spectral Range: 405 – 690 nm
Filter Slots: 6 wavelength
Precision: ± 0.01 OD at 0.000 – 0.500 OD
< 1% CV (< 2.0 OD)
< 1.5% CV (2.0 – 2.5 OD)
Accuracy: ± 0.01 OD or 2.5%
(0.000 – 3.000 OD) whichever is greater
Read Time: < 30 sec (single wavelength)
< 50 sec (dual wavelength)
Washer Specifications
Manifold Configuration: 8-channel wash head
Programmable Volumes: 50 – 999 μL
Wash Buffer Capacity: 5 wash bottles at 2.0 L, with level-sensing
Residual Wash Volume: < 3 μL per well with dual-axis sweep in a flat-bottom plate
Dispense Precision: ≤ 5% CV (with 300 μL in a 96 well plate)
Waste Container Capacity: 10 L, with level-sensing
Incubator Specifications
Number of Incubators: 12 (6 ambient, 6 elevated)
Temperature Range: RT + 4° – 45°C (elevated incubator)
RT + 4°C (ambient incubator)
Temperature Accuracy: ± 1°C
Shaking: 14 Hz (periodic or continuous)
Temperature Monitoring: Yes
Reagent Pipetting Specifications
Reagent Tip Size: 1,200 μL
Reagent Pipetting Volume: 20 – 1,000 μL
Reagent Pipetting Precision: ≤ 3% CV at 10 shots at any volume in operating range
Reagent Pipetting Accuracy: ± 2% of target volume (single-shot mode)
Sample Pipetting Specifications
Sample Tip Size: 300 µL
Sample Pipetting Volume: 10 – 300 µL
Sample Pipetting Precision: ≤ 3% CV at any operating volume above 10 µL
Sample Pipetting Accuracy: ± 2% of target volume
Time to Dispense: 11 minutes (typical)
50 µL of 96 samples to plate from sample tubes or deep-well plates
Dilution Range: 1 part in 199 one-stage dilution
1 part in 39,601 two-stage dilution
Process Security
Liquid Level Sensing: Yes (reagents, controls and samples)
Level Sensor System Pressure differential
Clot Detection: Yes
Dispense Anomaly Detection: Yes
Tip Detection: Yes
Electronic Signature Pipetting: Yes
Alarms: Yes
Related Products