RF-20A/RF-20Axs, Fluorescence Detector

Cat. No. Specification Unit Price Quantity
C002503359-01 Wavelength range: 0, 200 nm to 650 nm EA Inquiry
C002503359-02 Wavelength range: 0, 200 nm to 750 nm EA Inquiry

Add to Cart

Product Description
Product Description
Experience exquisite sensitivity with RF-20A and RF-20Axs fluorescence detectors for HPLC and UHPLC separations. Built with maintenance in mind, both the cell and lamp replacement are accessed through the front panel - no additional position adjustment is required. Monitor two EX and EM wavelength pairs simultaneously, or up to four pairs with LabSolutions software control. The RF-20xs has superior S/N performance due to its temerature controlled flow cell.
Features
• RF-20A has a wavelength range of 200 to 650 nm.
• RF-20AXS has a wavelength range of 200 to 750nm.
• Optional PMT or Photo-multiplier (both) to extend to 900nm
• Both detectors have wavelength accuracy of ±2 nm and precision of ±0.2nm.
• Superior with signal-to-noise ratio of at least 1200 for RF-20A and 2000 for RF-20Axs, allowing ultrafast analysis of trace-level analytes.
• Standard cell with a 12uL volume, or, a semi-micro cell with 3uL volume are available.
• Operating temperature 4 to 35°C.
Specifications
RF-20A
Light source: Xenon lamp
Wavelength range: 0, 200 nm to 650 nm
(Dual wavelength simultaneous monitoring possible)
Spectral bandwidth: 20nm
Wavelength accuracy: ±2 nm
Wavelength Reproducibility: ±0.2 nm
S/N: Water Raman peak S/N 1200 min.
Cell (capacity, pressure resistance, material): 12 µL; 2 MPa (approx. 20 kgf/cm2); SUS316L, PTFE (fluororesin), quartz

RF-20Axs
Light source: Xenon lamp, Low-pressure mercury lamp (to check wavelength accuracy)
Wavelength range: 0, 200 nm to 750 nm
(Dual wavelength simultaneous monitoring possible)
Spectral bandwidth: 20nm
Wavelength accuracy: ±2 nm
Wavelength Reproducibility: ±0.2 nm
S/N: Water Raman peak S/N 2000min.
Cell (capacity, pressure resistance, material): 12 µL; 2 MPa (approx. 20 kgf/cm2); SUS316L, PTFE (fluororesin), quartz
Cell Temperature Input Range: 4°C to 40°C, 1°C step
Cell Temperature Control Range: (Room temperature - 10°C) to 40°C
Related Products

Online Inquiry

Note: If you don't receive our verification email, do the following:

GET IN TOUCH
7x24

Seamless access to cutting-edge products and services with our professional support.

Copyright © Amerigo Scientific. All rights reserved.