HP Zero 1000 HC Tower Nitrogen Generator

Cat. No. Dimensions (WxDxH) Specification Unit Price Quantity
555102-01 175 x 490 x 670 mm (6.9 x 19.3 x 26.4 in) Flow rate: 1000 mL/min Each Inquiry

Add to Cart

Product Description
Product Description
The DBS® HP Tower Zero produces nitrogen by utilizing a combination of filtration and pressure swing adsorption (PSA) technology. Standard compressed air is filtered by high-efficiency coalescing filters to remove all contaminants down to 5 µm. For ultra-sensitive applications such as GC carrier and make-up gas, units also include the additional heated catalyst module to ensure hydrocarbons are removed to <0.1 ppm. The air then passes through two columns filled with a proprietary blended carbon molecular sieve (CMS) which adsorbs O2, CO2, and moisture. These are desorbed to the atmosphere during the pressure swing cycle leaving a supply of ultra-pure nitrogen.

· INCREASE EFFICIENCY
A constant gas supply with a guaranteed purity, eliminates interruptions of analyses to change cylinders and reduces the amount of instrument re-calibrations required.
· IMPROVE SAFETY
Nitrogen produced at low pressure and ambient temperature, removes the need for high-pressure cylinders.
· RETURN ON INVESTMENT
The payback period can be as short as 6 to 12 months.
· ENHANCE PERFORMANCE
Gas generators can be installed in the laboratory close to the instrument, eliminating the need for long gas lines from external cylinder supplies. A constant guaranteed high purity gas supply improves stability and ensures greater reproducibility of results.
Features
· Produces a continuous supply of high purity nitrogen
· On-demand supply 24/7
· Flow rate: 200 to 4000 mL/min
· Purity: up to +99.999% & <0.1 ppm THC
· Pressure: up to 5 barg (75 psig)
· Proprietary carbon molecular sieve technology
· 2-year complete product warranty
· Easy to install, operate and maintain

Benefits:
· Eliminates dangerous high-pressure cylinders helping to keep your employees safer
· Removes the logistics, inconvenience, downtime, and costs of a cylinder system
· Flow capacity to match your specific instrument demands
· Ideal for all GC applications - stable baseline with increased sensitivity and repeatability
· Meets and exceeds the requirements for the most demanding GC applications
· Superior air purification with long life catalyst technology
· Peace of mind
· Improve your laboratory workflow and productivity
Product Details
Certifications
CE, FCC
Specifications
Flow mL/min: 1000
Purity: >99.999%
Hydrocarbon purity (measured as methane): 0.1 ppm
Dewpoint °C (°F): -50 (-58)
Outlet pressure barg (psig): Up to 5 max (75)
Inlet pressure barg (psig): 7 to 10 (100 to 160)
Actual inlet air requirement liters @ 8 barg (116 psig): 17
Recommended compressor air inlet @ 8 barg (116 psig): 34
Pressure drop barg (psig): 1.5 (22)
Inlet air quality: Clean dry compressed air ISO8573-1:2010 Class 1.2.1
Technology: Carbon molecular sieve
Warm up time minutes: 60
LED indicators: Power on/off, system ready, errors
Electrical supply: 110-120V 60Hz / 220-240V 50 Hz
Power consumption watts: 270
Noise level: Minimal
Dimensions mm(in): 175W x 490H x 670D (6.9W x 19.3H x 26.4D)
Weight kg (lb): 26.9 (59.3)
Shipping dimensions mm (in): 770W x 410H x 590D (30.3W x 16.1H x 23.2D)
Shipping weight kg (lb): 31.9 (70)
Operating temp °C (°F): 15 to 35 (59 to 95)
Inlet connection: 1/4" Compression
Outlet connection: 1/8" Compression
Applications
Application Description
GC Applications
· GC-FID Carrier & Make-Up Gas
· ECD
· ELSD
· Head Space Vial Pressurization